Quality parameter of photon beam

During acceptance tests for the treatment unit depth-dose curves are measured (usually for fixed SSD).Parameter D10/D20 i.e. ratio of doses at depths 10 and 20 cm for field size 10cm x 10cm describes the beam quality.

Experimental data for SSD=100 cm

 

...and values as follows:

D10/20

MV

2.20

2.0

2.07

2.5

1.98

3.0

1.90

3.5

1.84

4.0

1.80

4.5

1.77

5.0

1.71

6.0

1.67

7.0

1.64

8.0

1.62

9.0

1.59

10.0

1.56

12.0

1.54

14.0

1.52

16.0

1.50

19.0

1.47

25.0

1.46

30.0

1.42

50.0